Laboratory evaluation of the CIP 10 personal dust sampler
Authors:A Gero  T Tomb
Institution:Mine Safety and Health Administration, Pittsburgh Health Technology Center, PA 15213.
Abstract:The "capteur individuel de poussiere" CIP 10 personal dust sampler--developed by the Centre d'Etudes et Recherches de Charbonnages de France (CERCHAR) research organization--is a small, quiet, lightweight unit which samples at a flow rate of 10 L/min. It is a three-stage sampler, using two stages to remove nonrespirable dust particles and one stage to collect the respirable fraction. Airflow through the sampler is induced by the third stage, which is a rotating collector cup that contains a fine grade sponge. Laboratory tests were conducted in a dust chamber using aerosols of Arizona road dust, coal dust and silica dust. Aerosol concentrations measured with the CIP 10 were compared to those measured with the coal mine dust personal sampler unit used in the United States. The results of this study showed that aerosol concentrations measured with the CIP 10 were linearly related to those obtained with the coal mine dust personal sampler. The relationship, however, was dependent on preselector configuration and aerosol characteristics. The collection medium allows some small particles (less than 3 microns) to pass through the sampler without being collected. As much as 13% (by weight) of the aerosol that penetrated through the preseparating stages was exhausted from the sampler.
Keywords: