首页 | 本学科首页   官方微博 | 高级检索  
     


Spectral reconstruction of high energy photon beams for kernel based dose calculations
Authors:Hinson William H  Bourland J Daniel
Affiliation:Department of Radiation Oncology, Wake Forest University School of Medicine, Winston-Salem, North Carolina 27157, USA. whinson@wfubmc.edu
Abstract:A kernel-based dose computation method with finite-size pencil beams (FSPBs) requires knowledge of the photon spectrum. Published methods of indirect spectral measurements using transmission measurements through beam attenuators use mathematical fits with a large number of parameters and constraints. In this study, we examine a simple strategy for fitting transmission data that models important physical characteristics of photon beams produced in clinical linear accelerators. The shape of an unattenuated bremsstrahlung spectrum is known, varying linearly from a maximum at zero energy to a value of zero at a maximum energy. This unattenuated spectrum is altered primarily by absorption of low energy photons by the flattening filter, causing the true spectrum to roll off to zero at low photon energies. A fitting equation models this behavior and has these advantages over previous methods: (1) the equation describes the shape of a bremsstrahlung spectrum based on physical expectations; and (2) only three fit parameters are required with a single constraint. Results for 4 MV and 6 MV accelerators for central axis and off-axis beams show good agreement with the maximum, average and modal energies for known spectra. Previously published models, representations of beam fluence (energy fluence, dN/dE), experimental methods, and the fitting process are discussed.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号