Needle core biopsy of the breast with a spring-loaded device. |
| |
Authors: | A J McMahon A M Lutfy A Matthew A D Walls J S McOrmick M A Henderson C D Auld |
| |
Affiliation: | Department of Surgery, Dumfries and Galloway Royal Infirmary, UK. |
| |
Abstract: | Over a 2-year period, 151 outpatients with a palpable breast lump requiring needle core biopsy were randomized to a Tru-Cut 14-G (T14), Bioptycut 14-G (B14) or Bioptycut 18-G (B18) needle. Use of a Biopty gun resulted in less pain than a Tru-Cut needle. An inadequate sample was obtained after two needle passes in 11 of 49 (T14), none of 51 (B14) and two of 51 (B18) patients (chi 2 = 14.6, 2 d.f., P = 0.0007). Tissue samples were assessed by a single pathologist for tissue volume and overall diagnostic value; the B14 group scored better than the B18 and T14 for both of these parameters (P < 0.003). The sensitivities were 68 (T14), 88 (B14) and 96 (B18) per cent (overall chi 2 = 7.3, 2 d.f., P = 0.026). The Biopty gun with a 14-G needle results in a higher sampling success rate, greater diagnostic sensitivity and a better specimen quality than the Tru-Cut, and is much easier to use. |
| |