螺旋CT低剂量扫描在新生儿腹部检查中的应用研究
引用本文:郭依廷,姚刚.螺旋CT低剂量扫描在新生儿腹部检查中的应用研究[J].中国中西医结合影像学杂志,2010,8(2):121-123.
作者姓名:郭依廷  姚刚
作者单位:1. 山东省枣庄市妇幼保健院,山东,枣庄,277102
2. 山东省枣庄市立医院,山东,枣庄,277102
摘    要:目的:评价新生儿腹部单排螺旋CT低剂量与常规剂量扫描图像质量,探讨低剂量扫描在新生儿应用的可行性。方法:对高度怀疑腹部病变的足月新生儿150例和早产儿90例,随机各分3组行常规(150mAs)及低剂量(40mAs、30mAs)腹部CT扫描,观察比较其对腹部正常结构的显示差异。结果:图像质量评价结果:足月新生儿150mAs,40mAs扫描获得的可诊断图像,χ2检验,P>0.05,提示图像无显著性差异;30mAs的可诊断图像,χ2检验,P<0.05,提示图像质量有显著性差异。早产儿150mAs,40mAs,30mAs可诊断图像,χ2检验,P>0.05,提示上述剂量图像质量差异无显著性意义。结论:单排螺旋CT低剂量扫描适用于新生儿腹部检查,在保证图像质量的前提下,足月新生儿采用40mAs、早产儿采用30mAs左右的扫描条件较为适宜且有利于患儿的辐射防护。

关 键 词:体层摄影术  X线计算机  腹部  低剂量  辐射防护

Application of low-dose single-slice spiral CT scanning in abdomen examination of newborn
GUO Yiting,YAO Gang.Application of low-dose single-slice spiral CT scanning in abdomen examination of newborn[J].Chinese Imaging Journal of Integrated Traditional and Western Medicine,2010,8(2):121-123.
Authors:GUO Yiting  YAO Gang
Institution:. (Maternal and Child Health Hospital of Zaozhuang, Zaozhuang, 277102, China)
Abstract:Objective: To evaluate the neonate abdominal images quality by single-slice spiral CT (SSCT) with low-dose and conventional parameters,and to study the feasibility of low-dose scan in the abdomen examination of newborn. Methods: One hundred and fifty cases mature neonates and 90 cases premature infants with suspected abdominal disease were selected and randomly divided into 3 groups ,then perform routine dose(150 mAs) and low dose(40 mAs,30 mAs)CT examination of newborn abdomen respectively, evaluated the presentation of the normal structure of each group. Results: Image quality evaluation: the percentage of the diagnostic image of mature neonate had no significant different between 150 mAs,40 mAs ( P 〉0.05) with x^2 test,and there was no significant different between 150 mAs, 40 mAs and 30 mAs images of premature infant. Conclusion: Low-dose SSCT scanning is suitable for abdomen examination of newborn. And it's suitable to choose tube current 40-30mAs in examination by SSCT for image quality assurance. And low-dose scanning also plays an important role in radioprotection.
Keywords:Tomography  X-ray computed  Abdomen  Low-dose  Radioproteetion