Reference intervals for the P-Albumin bromocresol purple method |
| |
Authors: | Peter Astrup Christensen |
| |
Affiliation: | Department of Clinical Biochemistry, Aalborg University Hospital, Aalborg, Denmark |
| |
Abstract: | Correct reference intervals are an important part of test results. As establishing own reference intervals is a very expensive task, the NORIP reference intervals are often transferred for use in Nordic laboratories. The NORIP reference interval on P-Albumin was here compared to current results for laboratories using the bromocresol purple (BCP) method for P-Albumin. External quality control reports were used to investigate the change in levels between the BCP and BCG methods on P-Albumin. An algorithm was built for extracting and isolating the laboratory’s healthy subject population. The algorithm was used to extract test results from the laboratory information system. Parametric and non-parametric statistical methods were used to evaluate the P-Albumin test result populations. The indirect method used here clearly shows that the NORIP reference intervals for P-Albumin are not fit for the current bromocresol purple methods. The method was also used to suggest new reference interval limits. |
| |
Keywords: | Albumin bromocresol BCG BCP green indirect purple reference interval |
|
|