首页 | 本学科首页   官方微博 | 高级检索  
     


The Superfast Atrial Recharge Pulse: A Cause of Pectoral Muscle Stimulation in Patients Equipped with a Unipolar DDD Pacemaker
Authors:ROLAND STROOBANDT,ROGER WILLEMS,PAUL DEPUYDT,&dagger  ,GERMAIN HOLVOET,ALFONS SINNAEVE
Affiliation:Department of Cardiology, St.-Jozef Hospital Oostende, Belgium.
Abstract:Pectoral muscle stimulation may cause serious discomfort to patients equipped with a pulse generator. Insulation defects of the lead, connector problems and defective coating of the pacemaker can are common causes of local muscle contractions. This report describes pectoral muscle stimulation caused by the atrial superfast recharge pulse incorporated into the atrial channel of a commercially available unipolar DDD pacemaker. As pectoral muscle stimulation could not be eliminated by reprogramming the pacemaker to a lower atrial output in some patients a redesign of the pacemaker is highly required.
Keywords:DDD pacemaker    muscle stimulation    atrial recharge pulse    blanking period
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号