首页 | 本学科首页   官方微博 | 高级检索  
检索        

超敏C反应蛋白试剂盒线性范围的验证
引用本文:赵彩虹.超敏C反应蛋白试剂盒线性范围的验证[J].国际检验医学杂志,2012,33(15):1819-1821.
作者姓名:赵彩虹
作者单位:北京市门头沟区医院检验科生化室,102300
摘    要:目的 验证本科室使用的超敏C反应蛋白(hs-CRP)试剂盒的线性范围,对该试剂盒的性能进行初步评价.方法 根据临床和实验室标准化协会(CLSI)制定的EP6-A2文件,配制6个试验样本,按随机顺序对它们进行测定,每个样本重复测定2次.经离群点检查后,计算比例误差(CVr)、多项式回归系数(bi)、残差(Bij)、标准误(Syx)、线性偏离(DLi).结果 所有测试结果未发现离群点,CVr为0.40%,小于试剂盒说明书的批内不精密度要求.二次回归多项式非线性系数b2、三次回归多项式非线性系数b3与0之间差异有统计学意义(P<0.05).二次回归多项式、三次回归多项式的Syx分别为0.029 3、0.018 7,三次回归多项式为最适多项式.6个试验样本的DLi最大值为1.22%,小于美国临床病理学家协会(CAP)要求的非线性差异要求(2.5%).结论 本科室使用的hs-CRP试剂盒在0~20.5 mg/L范围内呈线性,与试剂盒申明的线性范围相一致.

关 键 词:C反应蛋白质  线性模型  试剂盒  诊断

Demonstration of linear range of high-sensitive C-reactive protein assay kit
Zhao Caihong.Demonstration of linear range of high-sensitive C-reactive protein assay kit[J].International Journal of Laboratory Medicine,2012,33(15):1819-1821.
Authors:Zhao Caihong
Institution:Zhao Caihong(Biochemical Laboratory,Clinical Laboratory,Hospital of Mentougou District,Beijing 102300,China)
Abstract:Objective To demonstrate the linear range of high-sensitive C-reactive protein(hs-CRP) assay kit used in this laboratory,and evaluate the performance of this kit preliminarily.Methods According to the EP6-A2 document,described by Clinical and Laboratory Standards Institute(CLSI),6 samples were prepared,which were detected in duplicate in random order.After outlier test,the proportional error(CVr),polynomial regression coefficients(bi),residual(Bij),standard error(Syx) and deviation from linearity(DLi) were calculated.Results No outlier was found in all test results,and the CVr was 0.40%,less than the within-run imprecision requirements of kit instructions.There was a statistically significant difference between zero and nonlinear coefficient b2 in quadratic regression polynomial(P<0.05),so was b3 in cubic regression polynomial(P<0.05).The Syx of quadratic regression polynomial and cubic regression polynomial were 0.029 3 and 0.018 7 respectively,and the cubic regression polynomial was optimal.The maximum DLi was 1.22%,less than the nonlinear difference requirements of College of American Pathologists(CAP).Conclusion The hs-CRP assay kit used in this laboratory might be with linear range from 0 to 20.5 mg/L,in accordance with statement of kit instructions.
Keywords:C-reactive protein  linear models  reagent kits  diagnostic
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号