辐射剂量对头颅CT扫描图像质量的影响
引用本文:方先来,周全,刘斯润,张建针. 辐射剂量对头颅CT扫描图像质量的影响[J]. 放射学实践, 2001, 16(3): 205-207
作者姓名:方先来  周全  刘斯润  张建针
作者单位:1. 广东省粤北人民医院影像诊断科
2. 广州暨南大学附属第一医院医学影像诊断中心
3. 深圳市宝安区龙集医院CT室
摘    要:目的 :探讨不同辐射剂量行颅脑CT扫描对影像质量的影响。方法 :8例正常男性志愿者用 110~ 14 0kV ,180~ 36 0mAs分四组行轴位平扫 ;各影像按颗粒均匀性、脑灰白质细节、界面清晰度和有无射线硬化等积分评定质量。结果 :共扫描 192次 ,12 0kV、2 40mAs以上组影像质量均达优良 ;130kV、30 0mAs以上组优片率更高。结论 :采用适当辐射剂量 ( >12 0kV、2 40mAs)行颅脑CT扫描能提高密度分辨力 ,增加SNR ,改善影像质量

关 键 词:辐射剂量 头颅 CT 图像质量
文章编号:1000-0313(2001)03-0205-03

A study for imaging quality in the course of plain cranial CT scans with different radiation doses
Fang Xianlai,Zhou Quan,Zhang Jianzhen,et al.. A study for imaging quality in the course of plain cranial CT scans with different radiation doses[J]. Radiologic Practice, 2001, 16(3): 205-207
Authors:Fang Xianlai  Zhou Quan  Zhang Jianzhen  et al.
Affiliation:Fang Xianlai,Zhou Quan,Zhang Jianzhen,et al.Department of radiology,Yuebei People's Hospital,Shaoguan 512026
Abstract:Objective: To evaluate the effect of different radiation doses to imaging quality in the course of plain cranial CT scans.Methods:Eight normal male volunteers were divided into four groups,and each group was performed plain cranial axial CT scans with 110,120,130,140kV and 180 ~ 360mAs respectively. All images were scored at four degrees (excellent, good, poor, very poor) according to grain homogeneity, fine structure appearance of the grey and white matter, hardening of X rays and clearness of tissue interface. Results:The patients were totally scanned 192 times. The image quality with > 120kV and > 240mAs is excellent or good. The rate of superior image quality with ≥ 130kV and≥300mAs, moreover, was higher. Conclusion:Using radiation dose with > 120kV and > 240mAs, the density resolution and SNR may be increased and imaging quality be improved.
Keywords:Brain Tomography  X ray computed Image quality Evaluation