首页 | 本学科首页   官方微博 | 高级检索  
检索        


Transcatheter atrial septal defect closure: modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer septal occluder.
Authors:Karina M Carlson  Henri Justino  Richard E O'Brien  V Vivian Dimas  Glenn T Leonard  Ricardo H Pignatelli  Charles E Mullins  E O'Brian Smith  Ronald G Grifka
Institution:Lillie Frank Abercrombie Section of Cardiology, Texas Children's Hospital, Baylor College of Medicine, Houston, Texas 77030, USA.
Abstract:The objective of this study was to evaluate a new technique of sizing atrial septal defects (ASDs) for transcatheter device closure. ASD closure using the Amplatzer septal occluder (ASO) device is commonly performed. Complications, including arrhythmias, pericardial effusions, and perforations, may be related to oversizing ASDs and choosing larger devices. Two methods were used to size ASDs using a compliant balloon. In some patients, the balloon was inflated until a waist was visible (+)waist]; in others, only until no shunting was demonstrable by echocardiogram echo; (-)waist]. The device was selected and implanted using standard procedure and echo guidance. One hundred seventeen patients underwent secundum ASD closure with an ASO device. There were 43 patients in the (-)waist group and 74 in the (+)waist group. All devices were implanted successfully. The initial echo ASD diameter was larger in the (-)waist group compared to the (+)waist group (P = 0.01). There was a smaller difference between the initial echo and balloon-sized ASD diameters in the (-)waist group (P < 0.02). ASO device size implanted (in mm greater than echo ASD diameter) was smaller in the (-)waist group (P < 0.01). There were 0/43 complications in the (-)waist group and 5/74 in the (+)waist group. The complete closure rate was the same in both groups. Sizing an ASD by inflating a compliant balloon just until shunting is eliminated, and not until a waist is visible, results in less overstretching of the ASD and selection of a smaller ASO device, achieving similar closure rates and potentially fewer complications.
Keywords:atrial septal defect  balloon sizing  interventional catheterization
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号