首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of the Bactec 960 Automated Nonradiometric System for Isolation of Mycobacteria from Clinical Specimens
Authors:P. Rohner  B. Ninet  A.-M. Beni  R. Auckenthaler
Affiliation:Division des Maladies Infectieuses, Laboratoire Central de Bactériologie, H?pital Cantonal de Genève, 24 rue Micheli-du-Crest, 1211 Geneva 14, Switzerland e-mail: peter.rohner@hcuge.ch, CH
Abstract: The Bactec MGIT 960 system (Becton Dickinson, USA), designed for the culture of mycobacteria, was compared with the Bactec 460 instrument and culture on two solid egg-based media using a total of 1024 clinical specimens. Mycobacteria could be identified from 99 (9.7%) specimens, 89 (90%) of which were identified by the Bactec 960 system, 90 (91%) by the Bactec 460 system, and 82 (83%) by culture on the two egg-based media. The Bactec 960 cultures became positive an average of 16.7 days after specimen collection, the Bactec 460 cultures 14.9 days after collection, and the cultures on egg-based media 26.2 days after collection. The Bactec 960 is a compact and highly automated nonradiometric system that may replace the Bactec 460 system.
Keywords:
本文献已被 SpringerLink 等数据库收录!