螺旋式与环式椎体融合器的生物力学比较
引用本文:杨凯,王仁成,王信胜,燕好军,王俊勤,董军.螺旋式与环式椎体融合器的生物力学比较[J].中国临床解剖学杂志,2001,19(3):262-263.
作者姓名:杨凯  王仁成  王信胜  燕好军  王俊勤  董军
作者单位:泰山医学院附属医院骨科,
基金项目:国家留学基金资助项目(97837011)
摘    要:目的:用生物力学的方法比较螺旋式与环式椎体融合器的初始稳定性和拔出力量,方法:将24个颈椎运动节段(C4/5或C6/7),根据BMD和不同节段随机分为两组,分别用螺旋式BAK-C和环式Brantigan I/F,测试完整模型和固定状态下椎体在前屈/后伸,左右侧弯和轴向旋转6种运动状态的稳定性,最后进行拔出试验,结果:BAK-C和Brantigan I/F均能增加颈椎的初始稳定性,但在效率上无显著性差异,BAK-C的拔了出力量显著大于Brantigan I/F,结论:两种椎体融合器的稳定性几乎没有差异。螺旋式融合器具有较大的拔出力量。

关 键 词:颈椎  生物力学  椎体融合器  螺旋式  环式
文章编号:1001-165X(2001)03-0262-02
修稿时间:2000年5月24日

Biomechanical comparison of threaded and non-threaded interbody fusion devices
YANG Kai,WANG Ren cheng,WANG Xin sheng,et al..Biomechanical comparison of threaded and non-threaded interbody fusion devices[J].Chinese Journal of Clinical Anatomy,2001,19(3):262-263.
Authors:YANG Kai  WANG Ren cheng  WANG Xin sheng  
Institution:YANG Kai,WANG Ren cheng,WANG Xin sheng,et al. Department of Orthopedics,Affiliated Hospital of Taishan Medical College,Taian 271000,China
Abstract:Objective:To compare the initial stability and pullout strength of threaded and non threaded cervical interbody fusion devices on human specimens by biomechanical method.Methods:Flexibility of 24 cervical motion segments(two groups)were tested in axial rotation,flexion/extension and lateral bending,intact and after implantation of a cervical fusion device(threaded BKA C and non threaded Brantigan I/F).An implant pullout concluded testing.Changes in range of motion (ROM) were analyzed.Results:Both implants were effectvie at reducing ROM in all directions.Difference was found only just in the ROM of axial rotation.The pullout force of BAK C was significantly differenct between implants.Conclusion:Difference in initial stability between the two implants is rarely found.The threaded implant design has a higher pull out force.
Keywords:cervical spine  biomechanics  interbody fusion devices