EPID在颈、胸上段食管癌调强放疗中的应用价值
引用本文:胡俊,刘海,王晓萍,徐楠.EPID在颈、胸上段食管癌调强放疗中的应用价值[J].医疗设备信息,2012(6):164-166.
作者姓名:胡俊  刘海  王晓萍  徐楠
作者单位:解放军第八一医院全军肿瘤中心放疗科,江苏南京210002
摘    要:目的研究电子射野影像系统(EPID)在头颈肩面罩固定下颈、胸上段食管癌调强放疗中的效果。方法放疗前先采集定位片图像为参照,与治疗过程中实时采集的验证图像对比,测量摆位误差。结果入组30例患者共进行了120次治疗重复摆位,采集验证片120张比较每次摆位的体位误差。X轴(左右方向)3 mm以内误差有29例(96.67%),3~5 mm误差有1例(3.33%);Y轴(头足方向)3 mm以内误差有28例(93.33%),3~5 mm误差有2例(6.67%);Z轴(前后方向)3 mm以内误差有27例(90%),3~5 mm误差有3例(10%);X轴、Y轴、Z轴3个方向误差超过5 mm者未发现,我们对误差在3~5 mm患者进行重新摆位验证。结论利用电子射野影像系统可以有效减少放疗摆位误差,提高摆位的准确性,是质量控制和质量保证的有力工具。

关 键 词:调强放疗  电子射野影像系统  摆位误差  食管癌

The Application Value of EPID in Intensity-Modulated Radiotherapy for Neck and Upper Thoracic Esophageal Cancer
HU Jun,LIU Hai,WANG Xiao-ping,XU Nan.The Application Value of EPID in Intensity-Modulated Radiotherapy for Neck and Upper Thoracic Esophageal Cancer[J].Information of Medical Equipment,2012(6):164-166.
Authors:HU Jun  LIU Hai  WANG Xiao-ping  XU Nan
Institution:, Radiotherapy Department,Military Tumor Center,81st Hospital of PLA,Nanjing Jiangsu 210002,China
Abstract:Objective To study the efficacy of electronic portal imaging device(EPID) in intensity-modulated radiotherapy for neck and upper thoracic esophageal cancer,in which the patients were immobilized with the head/neck shoulder immobilization system.Methods The positioning images were collected before radiotherapy as reference images,which then can be compared to the verify images collected from real time,measuring set-up error.Results 30 patients enrolled a total of 120 times of repeated positioning during the treatments.120 verify films were collected.The set-up errors were analyzed by comparing the films of each positioning.X-axis(left and right direction) within 3 mm error is 29 patients(96.67%),3~5 mm error,1 patient(3.33%);Y-axis(head foot direction) within 3 mm error,28 patients(93.33%),3~5 mm error,2 patients(6.67%);Z axis(anterior-posterior direction) within 3 mm error,27 patients(90%),3~5 mm error in 3 patients(10%);no set-up error over 5mm was found in X axis,Y axis,and Z axis,re-positioning verification were made on positioning error over 3~5 mm.Conclusion The electronic portal imaging system,a powerful tool for quality control and quality assurance,it can effectively reduce the radiation set-up error and improve the positioning accuracy.
Keywords:intensity-modulated radiotherapy  electronic portal imaging system  set-up error  esophageal cancer