首页 | 本学科首页   官方微博 | 高级检索  
     


Complications of carotid artery stenting are largely preventable: a retrospective error analysis
Authors:Veeraswamy Ravi K  Rubin Brian G  Sanchez Luis A  Curi Michael A  Geraghty Patrick J  Parodi Juan C  Sicard Gregorio A
Affiliation:Division of Vascular Surgery, Department of Surgery at Emory University School of Medicine, Atlanta, GA 30322, USA. veeraswamy@emoryhealthcare.org
Abstract:Procedure-related complications with carotid artery stenting must be minimized for it to be a valid treatment for carotid stenosis. Failure analysis was done for 207 carotid stent procedures. All complications were reviewed and technical errors were identified. The procedure-related stroke rate was 2.9%, technical failure rate was 1.9%, and no patients died. Two strokes resulted from protocol deviations. A third stroke occurred while crossing a long, irregular lesion with a protection device instead of establishing flow reversal. Excessive instrumentation of the aortic arch resulted in 2 strokes. Distal embolization occurred after open-cell stenting a friable lesion in 1 patient. The majority of neurologic events and technical complications that occur during carotid stenting are preventable. By adhering to technical protocols, avoiding excessive instrumentation in the aortic arch, using flow reversal in selected lesions, and matching the appropriate anatomy and stent, results of carotid artery stenting can be exceptional.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号