首页 | 本学科首页   官方微博 | 高级检索  
检索        


Recording the oscillatory potentials of the electroretinogram with the DTL electrode
Authors:Pierre Lachapelle  Julie Benoit  John M Little  Benoît Lachapelle
Institution:(1) Department of Ophthalmology (D-289), McGill University-Montreal Children's Hospital, 2300 Tupper Street, H3H 1P3 Montreal, Quebec, Canada
Abstract:Suprathreshold photopic oscillatory potentials recorded with a DTL electrode were compared to those obtained with a Lovac corneal electrode. The overall oscillatory potential response (sum of oscillatory potentials) recorded with the DTL electrode was half of that obtained with the Lovac electrode. However, there was no evidence of a selective attenuation (or amplification) of any given oscillatory potential with the DTL electrode. Similarly, the oscillatory potential relative amplitude ratios and the peak times of the oscillatory potentials were identical for both electrodes. Our findings clearly indicate that the DTL electrode is adequate to record the high-frequency oscillatory potentials. Given the low cost and ease of use, as well as the disposable nature of the DTL electrode, we believe that electroretinographic specialists should seriously consider a wider utilization.
Keywords:DTL electrode  Electroretinogram  Oscillatory potentials  Technique
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号