首页 | 本学科首页   官方微博 | 高级检索  
检索        


Image Display for Collision Avoidance of Radiation Therapy: Treatment Planning
Authors:Max M Chao  Liang S Chao  Yu J Chen  Chuen M Hsieh  Shueh C Liou  Yuh L Lee  and Sang H Yen
Institution:Cancer Center and Cancer Research Group, Veterans General Hospital-Taipei, Taiwan, Republic of China.
Abstract:Patient treatment in a medical linear accelerator is characterized by many angular and translational movements of the gantry and couch. The direction and orientation of each treatment beam is specified by a set of gantry, turntable, and collimator angles. It is possible that some selected treatment field configurations will result in gantry/couch or gantry/patient collisions that remain undetected during the treatment planning process. In this work, a digital camera has been used to record all the workable gantry/ patient set-up images, and a Windows programming language is used to edit and display these images on a personal computer for the treatment planner to screen the treatment plans. These graphical displays enable the planner to be aware of any potential collision hazards by an actual visualization of each selected gantry/turntable or gantry/patient angle configuration.