自制氟碳声学造影剂对正常兔腹主动脉造影的实验研究
引用本文:马娜,任卫东,杨军,杨欣,唐力,喻晓娜.自制氟碳声学造影剂对正常兔腹主动脉造影的实验研究[J].中国超声医学杂志,2006,22(2):84-86.
作者姓名:马娜  任卫东  杨军  杨欣  唐力  喻晓娜
作者单位:110001,沈阳市,中国医科大学附属第一医院心脏血管检查中心
摘    要:目的 观察自制氟碳声学造影剂对正常兔腹主动脉的造影效果及不同造影剂注入方式造影的时间-强度曲线变化。 方法 经兔耳缘静脉注入0.5ml/kg自制造影剂,常规成像观察实时造影过程,间歇式触发成像定量分析团注和推注法(0.1ml/s)造影的时间-强度曲线变化。 结果 造影后腹主动脉灰阶及多普勒信号显著增强。团注法同推注方式比较,峰值强度显著增高,但曲线下降斜率快,平均通过时间短(P〈0.001);而推注方式可显著延长造影剂平均通过时间(P〈0.001)。 结论 自制氟碳声学造影剂能够显著增强血流显像,根据不同检查目的选择合适的造影剂注入方式,对提高造影效果有重要价值。

关 键 词:声学造影  腹主动脉  
收稿时间:2005-08-19
修稿时间:2005-12-01

Experimental Study on Enhanced Ultrasonography of Normal Abdominal Aorta using Self-made Perfluorocarbon Ultrasound Contrast Agent in Rabbits
Ma Na, Ren Weidong, Yang Jun, et al.Experimental Study on Enhanced Ultrasonography of Normal Abdominal Aorta using Self-made Perfluorocarbon Ultrasound Contrast Agent in Rabbits[J].Chinese Journal of Ultrasound in Medicine,2006,22(2):84-86.
Authors:Ma Na  Ren Weidong  Yang Jun  
Institution:Center of Cardiovascular Checking, the First Affiliated Hospital, China Medical University, Shenyang 110001 China
Abstract:Objective To observe the effect of enhanced ultrasonography on normal rabbit abdominal aorta using self-made perfluorocarbon ultrasound contrast agent and the time-intensity curve changes by different injection methods.Methods Self-made contrast agent was injected via rabbit ear vein at a dose of 0.5 m1/kg,real-time abdominal aorta enhancement was observed by conventional imaging,and changes of time-intensity curves were quantitatively analyzed using triggered imaging by bolus injection and infusion (0.1 m1/s).Results Gray-scale and Doppler signal of abdominal aorta were significantly enhanced by contrast agent.Peak intensity was much increased by bolus compared with by infusion and descending slope was rapid with mean transit time reduced(P<0.001),while mean transity time MTT was longer using infusion(P<0.001).Conclusions Self-made perfluorocarbon ultrasound contrast agent can significantly enhance blood flow display,it is importance to choice ideal injection method according to checking aim to improve the effect of enhanced ultrasonography.
Keywords:Enhanced ultrasonography  Abdominal aorta  Rabbit  