首页 | 本学科首页   官方微博 | 高级检索  
检索        


Surgical complications from hemostatic puncture closure devices
Authors:Eidt J F  Habibipour S  Saucedo J F  McKee J  Southern F  Barone G W  Talley J D  Moursi M
Institution:Department of Surgery, University of Arkansas for Medical Sciences and John L. McClellan VAMC, Little Rock, USA.
Abstract:BACKGROUND: For securing immediate hemostasis following percutaneous arterial catheterization, the Food and Drug Administration has approved three hemostatic puncture closure devices. We reviewed our institutional experience with one device (Angio-Seal). METHODS: A retrospective, single-center, nonrandomized observational study was made of all vascular complications following femoral cardiac catheterization. RESULTS: An immediate mechanical failure of the device was experienced in 34 (8%) patients. Surgical repair was required in 1.6% (7 of 425) of patients following Angio-Seal versus 0.3% (5 of 1662) following routine manual compression (P = 0.004). In 5 patients, the device caused either complete occlusion or stenosis of the femoral artery. The polymer anchor embolized in 1 patient and was retrieved with a balloon catheter at surgery. CONCLUSION: During the first year of utilization of a percutaneous hemostatic closure device following cardiac catheterization, we observed a marked increase in arterial occlusive complications requiring surgical repair. Surgeons must be familiar with the design of these devices to achieve precise repair of surgical complications.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏