Implantable cardioverter-defibrillator shocks increase T-wave alternans
Authors:Lampert Rachel  Soufer Robert  McPherson Craig A  Batsford William P  Tirado Stephanie  Earley Christine  Goldberg Anna  Shusterman Vladimir
Affiliation:From the Department of Medicine, Yale University School of Medicine New Haven, Connecticut, USA;;Department of Medicine, VA Connecticut Healthcare System, New Haven, Connecticut, USA;;and PinMed Inc., and Cardiovascular Institute, University of Pittsburgh, Pittsburgh, Pennsylvania, USA
Abstract:Introduction: While implantable defibrillator shocks save lives, shock can lead to ventricular arrhythmias. However, the mechanism of shock-related proarrhythmia remains unclear. We evaluated the impact of ICD shock on repolarization instability, a factor associated with ventricular arrhythmogenesis.
Methods and Results: Sixty-five patients with ICDs underwent ambulatory ECG monitoring during defibrillation testing 3 months postimplant. TWA was analyzed continuously in the time domain during baseline, sedated, and post-shock states. RR, QRS, and QT intervals and catecholamines were also measured continuously. Adequate pre- and post-shock Holter data were recorded in 55 patients, 48 male, mean 64 ± 12 years, 50 with coronary disease, 48 with prior spontaneous or induced arrhythmia. TWA significantly increased after shock, from 9.6 ± 0.5 to 11.9 ± 0.6 μV, as did QRS duration, epinephrine, and norepinephrine levels, compared with sedated and baseline states. RR intervals decreased minimally. TWA changes with shock were not associated with RR or QRS duration changes, but were associated with changes in epinephrine.
Conclusions: ICD shock, even in the sedated state, increases repolarization instability as measured by TWA, an effect mediated in part by sympathetic stimulation. This association between shock and TWA may have important mechanistic and clinical implications for optimization of defibrillation therapy.
Keywords:defibrillation    implantable cardioverter-defibrillator    T-wave alternans    repolarization
本文献已被 PubMed 等数据库收录!