Pentacam分析仪测量近视眼患者角膜厚度及影响因素分析
引用本文:张士胜,廖华萍,张静,姚文琴,戴南平,叶盛,董世奇,廉井财,王康孙.Pentacam分析仪测量近视眼患者角膜厚度及影响因素分析[J].眼科新进展,2012,32(1):44-47.
作者姓名:张士胜  廖华萍  张静  姚文琴  戴南平  叶盛  董世奇  廉井财  王康孙
作者单位:1. 200025 上海市,上海瑞视眼科;200025 上海市,上海交通大学医学院附属瑞金医院
2. 上海瑞视眼科, 上海市,200025
3. 上海交通大学医学院附属瑞金医院, 上海市,200025
基金项目:上海市重点学科建设项目(编号:S30205)~~
摘    要:目的使用Pentacam眼前节分析系统测量近视患者角膜中央最薄点厚度,并探讨与A型超声角膜测厚仪测量结果的差异及可能影响因素。方法对预行准分子激光角膜手术的近视患者150例(300眼),分别采用Pentacam系统及A型超声角膜测厚仪测量角膜中央最薄点厚度,对测量结果进行统计学分析。结果 Pentacam、A型超声角膜测厚仪测得的中央角膜最薄点厚度平均值分别为(543.7±26.3)μm、(539.2±27.4)μm,二者高度相关(r=0.97,P=0.0001)。Pentacam测得角膜厚度值偏厚,二者差值平均(4.4±7.0)μm,差异有统计学意义(t=10.94,P=0.0001),该差值与A型超声测量角膜厚度值、年龄有关(P=0.0001),而与屈光度、性别、角膜直径、角膜曲率等无关。结论 Pentacam测量近视患者角膜厚度值略高于A型超声角膜测厚仪,尽管二者有很好的协同性,但相互间还不能完全替代。

关 键 词:Pentacam系统  超声测厚仪  角膜厚度  近视

Corneal thickness measured by Pentacam in myopic eyes and its related factors analysis
ZHANG Shi-Sheng , LIAO Hua-Ping , ZHANG Jing , YAO Wen-Qin , DAI Nan-Ping , YE Sheng , DONG Shi-Qi , LIAN Jing-Cai , WANG Kang-Sun.Corneal thickness measured by Pentacam in myopic eyes and its related factors analysis[J].Recent Advances in Ophthalmology,2012,32(1):44-47.
Authors:ZHANG Shi-Sheng  LIAO Hua-Ping  ZHANG Jing  YAO Wen-Qin  DAI Nan-Ping  YE Sheng  DONG Shi-Qi  LIAN Jing-Cai  WANG Kang-Sun
Institution:From the New Vision Eye Clinic(ZHANG Shi-Sheng,LIAO Hua-Ping,ZHANG Jing,YAO Wen-Qin,DAI Nan-Ping,YE Sheng,DONG Shi-Qi,WANG Kang-Sun),Shanghai 200025,China;Department of Ophthalmology,Ruijin Hospital,School of Medicine,Shanghai Jiaotong University(ZHANG Shi-Sheng,LIAN Jing-Cai,WANG Kang-Sun),Shanghai 200025,China
Abstract:Objective To compare the difference of the thinnest central corneal thickness in myopic eyes using Pentacam and A-scan ultrasound pachymetry,and assess the related effective factors.Methods The thinnest central corneal thickness in 300 eyes of 150 myopic patients before laser refractive surgeries was measured by Pentacam and A-scan ultrasonic pachymetry.The differences in the measurements obtained by two methods were statistically analyzed.Results The thinnest central corneal thickness measured by Pentacam and A-scan ultrasonic pachymetry were(543.7±26.3)μm and(539.2±27.4)μm,respectively,the correlation between them was high(r=0.97,P=0.000 1).The thickness measured by Pentacam was thicker than that by A-scan ultrasonic pachymetry,the difference was(4.4±7.0)μm,there was significant difference(P=0.000 1).The difference was correlated with the corneal thickness measured by A-scan ultrasonic pachymetry and the age(P=0.000 1),while there was no correlation with refractive power,gender,corneal diameter and corneal curvature.Conclusion The corneal thickness measured by Pentacam is thicker than that by A-scan ultrasound pachymetry in myopic eyes.Although they have well cooperativity,they cannot be simply used interchangeably.
Keywords:Pentacam  ultrasonic pachymetry  corneal thickness  myopia