Vibration sensitivity as a parameter for detecting peripheral neuropathy |
| |
Authors: | Hans Muijser Jacob Hooisma Elisabeth M G Hoogendijk Divera A M Twisk |
| |
Institution: | (1) Medical Biological Laboratory TNO, P.O. Box 45, 2280 AA Rijswijk, The Netherlands |
| |
Abstract: | Summary Two methods of evaluating the threshold for vibration perception were compared. Surprisingly it appears that the theoretically attractive, adaptive forced choice method does not result in lower variability than the method of limits. Moreover two devices were used to evaluate the threshold: the Optacon Tactile Tester and the multirod . Based on the characteristics of these devices and the known properties of mechanoreceptors, it is argued that the two devices test different mechanoreceptor systems. The high correlation of threshold with age (r = 0.9) found by Arezzo and Schaumburg 1] in measurements with the Optacon could not be reproduced. |
| |
Keywords: | Forced choice Meissner receptors Optacon Pacini receptors Peripheral neuropathy |