Reproducibility of left ventricular volume determinations with use of a semi-automated system |
| |
Authors: | Rolf Loertscher Prof. Felix Burkart Hans Erich Schmitt Hans Emmenegger |
| |
Affiliation: | (1) Division of Cardiology and Cardiovascular Radiology, Department of Internal Medicine and Radiology, University Hospital, CH-4031 Basel, Switzerland |
| |
Abstract: | Reliability of a computer-assisted system for determination of left ventricular volumes was judged by multiple measurements of rotation ellipsoids, cadaver hearts, and cineangiograms from patients. The volume measurements in cadaver hearts provided a volume correction factor necessary for reproducible results. Variation coefficient for intraobserver and interobserver variability did not exceed 2.3% when calculated using rotation ellipsoids and was highest at 12.0% for the end-systolic volumes derived from patient films. When appropriate calibration methods are employed, different observers can make reliable left ventricular volume measurements aided by such systems. |
| |
Keywords: | Heart, volume measurement, observer variability Cineangiography, biplane Ventricular volume determination Heart, volume correction factor Computer-assisted volume measurement system |