首页 | 本学科首页   官方微博 | 高级检索  
     


Reproducibility of left ventricular volume determinations with use of a semi-automated system
Authors:Rolf Loertscher  Prof. Felix Burkart  Hans Erich Schmitt  Hans Emmenegger
Affiliation:(1) Division of Cardiology and Cardiovascular Radiology, Department of Internal Medicine and Radiology, University Hospital, CH-4031 Basel, Switzerland
Abstract:Reliability of a computer-assisted system for determination of left ventricular volumes was judged by multiple measurements of rotation ellipsoids, cadaver hearts, and cineangiograms from patients. The volume measurements in cadaver hearts provided a volume correction factor necessary for reproducible results. Variation coefficient for intraobserver and interobserver variability did not exceed 2.3% when calculated using rotation ellipsoids and was highest at 12.0% for the end-systolic volumes derived from patient films. When appropriate calibration methods are employed, different observers can make reliable left ventricular volume measurements aided by such systems.
Keywords:Heart, volume measurement, observer variability  Cineangiography, biplane  Ventricular volume determination  Heart, volume correction factor  Computer-assisted volume measurement system
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏