首页 | 本学科首页   官方微博 | 高级检索  
检索        


Asymmetry of Doppler spectrum in stenosis differentiation
Authors:K Kaluzynski  A Tedgui
Institution:1. Biomedical Engineering Group, IBSPiE, Warsaw Technical University, ul. Chodkiewicza 8, 02-525, Warsaw, Poland
2. Institut National de la Santé et de la Recherche Médicale, Unité 141, H?pital Lariboisière, 41 bd de la Chapelle, 75475, Paris Cedex 10, France
Abstract:The asymmetry of the spectral distribution of ultrasonic Doppler flow velocity signals, assessed using the coefficient of skewness, is discussed as a criterion of stenosis differentiation. Its performance is compared with that of the index of turbulence intensity for both in vitro and in vivo flow Doppler signals, recorded distal to a stenosis. The power spectral distributions are computed using the direct Fourier transform and maximum likelihood method. The asymmetry of spectral distribution has proved to be a more efficient criterion than the turbulence intensity. The maximum likelihood method ensures better stenosis differentiation than the direct FFT method.
Keywords:Asymmetry of spectrum  Fast Fourier transform  Maximum likelihood method  Spectrum analysis  Stenosis differentiation  Turbulence intensity  Ultrasonic Doppler signal
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号