Mechanical Characterization of an In Vitro Device Designed to Quantitatively Injure Living Brain Tissue |
| |
Authors: | Barclay Morrison III David F. Meaney Tracy K. McIntosh |
| |
Affiliation: | (1) Department of Bioengineering, University of Pennsylvania, Philadelphia, PA;(2) Department of Neurosurgery, University of Pennsylvania, Philadelphia, PA |
| |
Abstract: | Due to the nonlinear, viscoelastic material properties of brain, its mechanical response is dependent upon its total strain history. Therefore, a low strain rate, large strain will likely produce a tissue injury unique from that due to a high strain rate, moderate strain. Due to a lack of current understanding of specific in vivo physiological injury mechanisms, a priori assumptions cannot be made that a low strain rate injury induced by currently employed in vitro injury devices is representative of clinical, nonimpact, inertial head injuries. In the present study, an in vitro system capable of mechanically injuring cultured tissue at high strain rates was designed and characterized. The design of the device was based upon existing systems in which a clamped membrane, on which cells have been cultured, is deformed. However, the present system incorporates three substantial improvements: (1) noncontact measurement of the membrane deflection during injury; (2) precise and independent control over several characteristics of the deflection; and (3) generation of mechanical insults over a wide range of strains (up to 0.65) and strain rates (up to 15s–1). Such a system will be valuable in the elucidation of the mechanisms of mechanical trauma and determination of injury tolerance criteria on a cellular level utilizing appropriate mechanical injury parameters. |
| |
Keywords: | Cell culture Mechanical stimulus Stretch injury Dynamic loading Device Brain tissue Brain injury |
本文献已被 SpringerLink 等数据库收录! |
|