首页 | 本学科首页   官方微博 | 高级检索  
     


Estimation of Radiation Dose in CT Based on Projection Data
Authors:Xiaoyu Tian  Zhye Yin  Bruno De Man  Ehsan Samei
Affiliation:1.Department of Biomedical Engineering, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology,Duke University Medical Center,Durham,USA;2.CT Systems and Applications Laboratory, GE Global Research,Niskayuna,USA;3.Medical Physics Graduate Program, Departments of Radiology, Electrical and Computer Engineering, and Physics,Duke University Medical Center,Durham,USA
Abstract:Managing and optimizing radiation dose has become a core problem for the CT community. As a fundamental step for dose optimization, accurate and computationally efficient dose estimates are crucial. The purpose of this study was to devise a computationally efficient projection-based dose metric. The absorbed energy and object mass were individually modeled using the projection data. The absorbed energy was estimated using the difference between intensity of the primary photon and the exit photon. The mass was estimated using the volume under the attenuation profile. The feasibility of the approach was evaluated across phantoms with a broad size range, various kVp settings, and two bowtie filters, using a simulation tool, the Computer Assisted Tomography SIMulator (CATSIM) software. The accuracy of projection-based dose estimation was validated against Monte Carlo (MC) simulations. The relationship between projection-based dose metric and MC dose estimate was evaluated using regression models. The projection-based dose metric showed a strong correlation with Monte Carlo dose estimates (R 2?>?0.94). The prediction errors for the projection-based dose metric were all below 15 %. This study demonstrated the feasibility of computationally efficient dose estimation requiring only the projection data.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号