TOSOH HLC-723 G8全自动糖化血红蛋白分析仪性能评价
引用本文:陈健. TOSOH HLC-723 G8全自动糖化血红蛋白分析仪性能评价[J]. 检验医学, 2013, 0(10): 925-927
作者姓名:陈健
作者单位:复旦大学附属华山医院检验科,上海200040
摘    要:目的对以离子交换高效液相色谱法(HPLC)为原理的TOSOHHLC-723G8全自动糖化血红蛋白分析仪(简称HLC-723G8)检测全血糖化血红蛋白(HbA1c)的性能进行评价。方法对HLC-723G8测定HbA1c的精密度、相关性、线性、携带污染进行评价,并检测54例健康体检者的HbA1c进行参考区间验证。结果HLC-723G8测定HbA1c低、高水平批内变异系数(cv)分别为0.00%、0.47%,批间CV分别为0.76%、0.31%;HLC-723G8与HLC-723G7全自动糖化血红蛋白分析仪测定HbA1c的结果呈明显相关(F2=0.999,P〈0.001),预期偏倚为0.06,相对偏倚值为0.83%,偏差符合率为100%;在5%~12%范围内线性良好(r2=0.9996,P〈0.001);高值标本对低值标本的结果无明显携带污染(携带污染率为1.12%);54名健康人群的HbA1c测定结果为4.7%-6.3%,在厂家提供的参考区间内。结论HLC-723G8测定HbA1c的性能良好,可在临床使用。

关 键 词:高效液相色谱法  糖化血红蛋白  性能评价

Performance evaluation of TOSOH HLC-723 G8 automatic glycohemoglobin analyzer
CHEN Jian. Performance evaluation of TOSOH HLC-723 G8 automatic glycohemoglobin analyzer[J]. Laboratory Medicine, 2013, 0(10): 925-927
Authors:CHEN Jian
Affiliation:CHEN Jian( Department of Clinical Laboratory, Huashan Hospital, Fudan University, Shanghai 200040, China)
Abstract:Objective To evaluate the performance of TOSOH HLC-723 G8 automatic glycohemoglobin analyzer (HLC-723 G8) for the determination of glycosylated hemoglobin A1c (HbA1c) with the principle of ion-exchange high- performance liquid chromatography (HPLC). Methods The precision, correlation, linearity and carryover of HLC-723 G8 were evaluated. The reference interval of HbAlc was verified among 54 healthy subjects. Results The within-run and between-run coefficients of variation (CV) of both low and high levels of HbA1c were 0.00% ,0.47% and 0.76% , 0.31%, respectively. HLC-723 G8 and HLC-723 G7 for the determination of HbAjc showed good correlation ( r2 = 0. 999 ,P 〈0.001 ). The expected bias was 0.06 ,the relative bias was 0.83% ,and the bias coincidence was 100%. An excellent linearity was found for HbA1c in the range of 5% - 12% (r2 = 0. 999 6 ,P 〈 0. 001 ). There was no obvious carryover in both low and high level samples, and the carryover rate was 1.12%. The levels of HbA1c among the 54 healthy subjects were 4.7% -6.3% in the range provided by manufacturers. Conclusions HLC-723 G8 has a good performance, and can be used in clinical application.
Keywords:High-performance liquid chromatography  Glycosylated hemoglobin Ale  Performance evaluation