彩色多普勒血流显像检测下肢静脉瓣返流
引用本文:查长松,赵玉华,杜晓华,李静伟,隋向梅.彩色多普勒血流显像检测下肢静脉瓣返流[J].中国医学影像技术,1998,14(12):922-924.
作者姓名:查长松  赵玉华  杜晓华  李静伟  隋向梅
作者单位:1. 山东莱阳解放军第145医院特检科 265200
2. 上海市第二军医大学长海医院
3. 海军412医院(进修生)
摘    要:目的:为探讨彩色多普勒血流显像(CDFI)检测静脉瓣返流的准确性。方法:40条正常和30条慢性静脉瓣功能不全(CVI)下肢进行了CDFI检查,将下肢静脉分为股总静脉(CFV)、股浅静脉(SFV)、静脉(PPV)、胫后静脉(PTV)和隐股连接部(SFJ)五个节段。结果:正常下肢480个静脉节段的生理性返流时间95%正常参考值上限为0.5s。以静脉造影为对照,患肢异常节段较正常节段返流时间显著延长(P<0.01),以返流时间>0.5s做为静脉瓣病理返流的标准,立位乏氏法和挤压法的敏感性分别为81.6%、85.7%,特异性100%、90.9%和准确性85.0%、86.7%。结论:本文提示CDFI是无创性评价静脉瓣功能不全敏感且准确的方法。

关 键 词:慢性静脉瓣功能不全  静脉瓣返流  彩色多普勒血流显像  静脉造影
收稿时间:8/2/1998 12:00:00 AM
修稿时间:1998年8月13日

Venous Valvular Reflux in Lower LimBs Assessed By Color Doppler Flow Imaging
Zha Changsong,Zhao Yuhu,Du Xiaohua.Venous Valvular Reflux in Lower LimBs Assessed By Color Doppler Flow Imaging[J].Chinese Journal of Medical Imaging Technology,1998,14(12):922-924.
Authors:Zha Changsong  Zhao Yuhu  Du Xiaohua
Institution:Laiyang 265200
Abstract:Purpose:To explore the accuracy of venous valvular reflux assessed By CDFI.Methods:30 lower limB veins with CVI and 40 normal limB veins were studied.The lower limB veins were divided into 5 segments:CFV,SFV,PPV,PTV and SFJ.The CVI limBs were also examined By phyleBography.Result:Physiological valvular reflux times of 480 normal segments were 0.23±0.14 sec,the 95% normal upper limit value was 0.5 sec.The reflux times in CVI limB reflux segments confirmed By phyleBography were significantly longer than that in normal segment( P <0.01).The reflux time >0.5 sec as the standard for diagnosing venous valvular reflux,the sensitivity for erect Valsalva and limB compression maneuver were 81.6%,85.7%, specificity 100%,90.9% and accuracy 85.0%,86.7% respectively.Conclusion:It is suggested that CDFI were noninvasive and valuaBle and practical for assessing venous function in lower limBs.