电离辐射诱导小鼠骨电子顺磁共振信号检测的影响因素
引用本文:任越,张慧芳,胡波,王婧洁,王宏利,原雅艺,张睿凤,李幼忱. 电离辐射诱导小鼠骨电子顺磁共振信号检测的影响因素[J]. 国际放射医学核医学杂志, 2015, 39(5): 371-374. DOI: 10.3760/cma.j.issn.1673-4114.2015.05.004
作者姓名:任越  张慧芳  胡波  王婧洁  王宏利  原雅艺  张睿凤  李幼忱
作者单位:030006 太原,中国辐射防护研究院放射医学与环境医学研究所
基金项目:中国辐射防护研究院院青年基金项目YQ130202
摘    要:目的 研究影响小鼠骨辐射诱导电子顺磁共振(EPR)信号检测的因素。 方法 将小鼠股骨去除软组织和骨髓,干燥,制成股骨段,经60Co γ射线照射,照射剂量率为5 Gy/min,照射剂量为10 Gy和50 Gy。检测不同干燥程度、检测方向、微波功率和叠加次数等条件下骨EPR的信号强度。 结果 降低骨样品含水量、设置合适的微波功率、增加叠加次数均可以增加样品信号强度,且不同检测方向下同一份样品的信号强度有一定差异。 结论 适当的样品预处理和检测参数设定可以提高骨EPR检测的灵敏度,样品多方向检测并取平均值可减弱骨样品不均一产生的影响。

关 键 词:电离辐射     电子顺磁共振   小鼠
收稿时间:2015-03-16

Influencing factors of detection on ionizing radiation-induced electron paramagnetic resonance signal in bones
Yue Ren,Huifang Zhang,Bo Hu,Jingjie Wang,Hongli Wang,Yayi Yuan,Ruifeng Zhang,Youchen Li. Influencing factors of detection on ionizing radiation-induced electron paramagnetic resonance signal in bones[J]. International Journal of Radiation Medicine and Nuclear Medicine, 2015, 39(5): 371-374. DOI: 10.3760/cma.j.issn.1673-4114.2015.05.004
Authors:Yue Ren  Huifang Zhang  Bo Hu  Jingjie Wang  Hongli Wang  Yayi Yuan  Ruifeng Zhang  Youchen Li
Affiliation:Department of Radiological Medicine and Environmental Medicine, China Institute for Radiation Protection, Taiyuan 030006, China
Abstract: Objective To study the influencing factors of detection on ionizing radiation-induced electron paramagnetic resonance(EPR) signal in mice bone. Methods Soft tissue and bone marrow samples of mice femur were collected, and femur sections were dried. The femur sections were irradiated at 10 and 50 Gy by 60Co sources with irradiation dose rate of 5 Gy/min. Bone EPR signal intensity was detected under different degrees of drying, detection direction, and microwave power, as well as various scan times. Results The water content of the bone sample was reduced, and microwave power was set to an appropriate value. Results showed that increasing scan times increased the signal intensity. Additionally, different detection directions of the same sample resulted in differences in signal intensity. Conclusion Appropriate sample preparation and detection parameters can improve bone EPR detection sensitivity. Multidirectional detection of samples and averaging of the results may reduce the effects of inhomogeneity on bone samples.
Keywords:Ionizing radiation  Bone  Electron paramagnetic resonance  Mice