A dosimetric and centeredness comparison of the conventional and novel endobronchial applicators: A preliminary study
Authors:Kotaro Iijima  Naoya Murakami  Hiroyuki Okamoto  Satoshi Nakamura  Shie Nishioka  Takahito Chiba  Junichi Kuwahara  Hiroki Nakayama  Mihiro Takemori  Ako Aikawa  Shibata Yoshihiro  Tairo Kashihara  Kana Takahashi  Koji Inaba  Kae Okuma  Hiroshi Igaki  Yuko Nakayama  Jun Itami
Affiliation:1. Department of Medical Physics, National Cancer Center Hospital, Chuo City, Tokyo, Japan;2. Department of Radiation Oncology, National Cancer Center Hospital, Chuo City, Tokyo, Japan;3. Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan;4. Department of Radiological Technology, National Cancer Center Hospital, Chuo City, Tokyo, Japan
Abstract:PurposeThis study compared the applicator position relative to the tracheal wall and dosimetric parameters between conventional and novel applicators among patients receiving endobronchial brachytherapy (EBBT) for intratracheal tumors.Methods and MaterialsData from 7 patients who received EBBT for intratracheal tumors were retrospectively analyzed; 4 and 3 patients were treated with conventional (2-wing) or novel (5-wing) applicators, respectively. Applicator centrality was evaluated using the distance between the center of the trachea and main bronchus (TMB) lumen and path of source (L). Dosimetric parameters, including plans normalized to D2cc of the TMB = 45 Gy (normalized plan), were compared between the applicators.ResultsThe mean and maximum values of L in cases of the 2-wing applicator group were approximately 5.0 mm and 10.0 mm, respectively. In the novel applicator group, the corresponding values were approximately 3.0 and 6.0 mm, respectively. In the normalized plan of the 2-wing applicator group, the ranges of median V90% of clinical target volume (CTV) and D0.1cc of the TMB in all cases were 23.0–91.9% and 66.3–153.1 Gy, respectively. In the 5-wing applicator group, the corresponding values were 69.2–83.8% and 60.4–84.5 Gy, respectively.ConclusionsIn the 5-wing applicator group, the range was narrow in all dose–volume parameters except for D2cc of the TMB. Compared to the conventional applicator, the 5-wing applicator can give a stable dose to the CTV and can reduce the maximum dose of the TMB. This suggests that stable EBBT can be given to any patient using the 5-wing applicator.
Keywords:Endobronchial brachytherapy  A novel applicator  Dosimetric comparison
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号