Inverse Planning – a Comparative Intersystem and Interpatient Constraint Study
Authors:Dietmar Georg PhD  Bernhard Kroupa  Petra Georg  Peter Winkler  Joachim Bogner  Karin Dieckmann  Richard Pötter
Institution:Department of Radiotherapy and Radiobiology, AKH Vienna, Medical University of Vienna, Vienna, Austria. Dietmar.Georg@akhwien.at
Abstract:PURPOSE: To compare commercial treatment-planning systems (TPS) for inverse planning (IP) and to assess constraint variations for specific IMRT indications. MATERIAL AND METHODS: For IP, OTP, XiO and BrainSCAN were used and step-and-shoot intensity-modulated radiotherapy (IMRT) delivery was assumed. Based on identical constraints, IP was performed for a prostate, head and neck, brain, and gynecologic case. IMRT plans were compared in terms of conformity/homogeneity, dose-volume histograms (DVHs), and delivery efficiency. For ten patients each of a class of indications, constraint variations were evaluated. RESULTS: IMRT plans were comparable concerning minimum target dose, homogeneity, conformity, and maximum doses to organs at risk. Larger differences were seen in dose gradients outside the target, monitor units, and segment number. Using help structures proved efficient to shape isodoses and to reduce segmentation workload. For IMRT class solutions, IP constraint variations depended on anatomic site. CONCLUSION: IP systems requiring doses as input and having objective functions based on physical parameters had a very similar performance. Constraint templates can be established for a class of IMRT indications.