首页 | 本学科首页   官方微博 | 高级检索  
检索        

两种不同检测系统测定血清总蛋白的可比性分析
引用本文:谭柏松,刘光明,张伟坚,陈世豪,李珍瑶.两种不同检测系统测定血清总蛋白的可比性分析[J].检验医学与临床,2013,10(6):689-690,693.
作者姓名:谭柏松  刘光明  张伟坚  陈世豪  李珍瑶
作者单位:南方医科大学附属新会医院检验科,广东江门,529100
摘    要:目的探讨COBAS C501模块化全自动生化分析仪与Biosystems A15全自动蛋白分析仪检测结果是否具有可比性。方法依据国家临床实验室标准委员会(NCCLS)EP9-A2文件要求,分别用两种检测系统测定血清总蛋白(TP),计算实验检测系统(Y)和目标检测系统(X)之间的系统误差,以美国临床医学检验部门修正法(CLIA′88)建议的医学决定水平处的系统误差来判断不同检测系统之间的可比性。结果在所测定TP中,预期偏差在方法线性范围内均可以被接受。结论 TP可在两检测系统上任意检测。当同一项目在不同仪器上检测时,结果应进行比对和偏倚评估,以保证检测结果的可比性。

关 键 词:血清总蛋白  比对  偏倚评估

Comparative analysis on detection results of serum total protein between two different detection systems
TAN Bai-song , LIU Guang-ming , ZHANG Wei-jian , CHEN Shi-hao , LI Zhen-yao.Comparative analysis on detection results of serum total protein between two different detection systems[J].Laboratory Medicine and Clinic,2013,10(6):689-690,693.
Authors:TAN Bai-song  LIU Guang-ming  ZHANG Wei-jian  CHEN Shi-hao  LI Zhen-yao
Institution:(Department of Clinical Laboratory ,Af- filiated Xinhui Hospital, Southern Medical University, Jiangmen, Guangdong 529100, China)
Abstract:Objective To explore the comparability of detection results by COBAS C501 and Biosystems A15 detection systems. Methods According to National Committee for Clinical Laboratory Standards (NCCLS) document EPgA2,we tested total protein(TP) with two detection systems. Then the system bias between different test sys tems was evaluated by obtained equation. The comparability of different investigated system was judged according to the total error allowed by clinical laboratory improvement amendment 88(CLIA'88). Results Among these results of TP,the anticipated bias was acceptable within methodological linear limitation. Conclusion TP could be tested by either of the two analytical systems. But for the tests, there are more than two sets of analytical systems in the labora- tory, the analytical comparison and bias evaluation should be made. Only after accurate evaluation evidence about the bias of their items in the different analytical systems is obtained, the accuracy and stability of testing results could be ensured.
Keywords:total protein  comparison  bias evaluation
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号