Percutaneous Venous Thrombectomy Using the Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) with Temporary Caval Filtration: In Vitro Investigations
Authors:Joachim Ernst Wildberger  Patrick Haage  Jan Bovelander  Joachim Pfeffer  Claudia Weiss  Dierk Vorwerk  Thomas Schmitz-Rode  Rolf W. Günther
Affiliation:(1) Department of Diagnostic Radiology, University of Technology, Aachen, Germany;(2) Department of Clinical Research, Arrow International, Delft, The Netherlands;(3) Department of Medical Statistics, University of Technology, Aachen, Germany;(4) Department of Diagnostic and Interventional Radiology, Klinikum Ingolstadt, Ingolstadt, Germany;(5) Department of Diagnostic Radiology, University Hospital, RWTH Aachen, Pauwelsstrasse 30, Aachen, 52074, Germany
Abstract:Purpose To evaluate the size and quantity of downstream emboli after thrombectomy using the Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) with or without temporary filtration for extensive iliofemoral and iliocaval thrombi in an in vitro flow model.Methods Iliocaval thrombi were simulated by clotted bovine blood in a flow model (semilucent silicone tubings, diameter 12–16 mm). Five experimental set-ups were performed 10 times each; thrombus particles and distribution were measured in the effluent. First, after retrograde insertion, mechanical thrombectomy was performed using the PTD alone. Then a modified self-expanding tulip-shaped temporary vena cava stent filter was inserted additionally at the beginning of each declotting procedure and removed immediately after the intervention without any manipulation within or at the filter itself. In a third step, the filter was filled with thrombus only. Here, two experiments were performed: Careful closure within the flow circuit without any additional fragmentation procedure and running the PTD within the filter lumen, respectively. In the final set-up, mechanical thrombectomy was performed within the thrombus-filled tubing as well as in the filter lumen. The latter was closed at the end of the procedure and both devices were removed from the flow circuit.Results Running the PTD in the flow circuit without filter protection led to a fragmentation of 67.9% (±7.14%) of the clot into particles le500 mgrm; restoration of flow was established in all cases. Additional placement of the filter safely allowed maceration of 82.9% (±5.59%) of the thrombus. Controlled closure of the thrombus-filled filter within the flow circuit without additional mechanical treatment broke up 75.2% (±10.49%), while additional mechanical thrombectomy by running the PTD within the occluded filter led to dissolution of 90.4% (±3.99%) of the initial clot. In the final set-up, an overall fragmentation rate of 99.6% (±0.44%) was achieved.Conclusions The combined use of the Arrow-Trerotola PTD and a temporary vena cava stent filter proved to be effective for even large clot removal in this experimental set-up.
Keywords:Catheters and catheterization  Fragmentation, mechanical  Interventional procedures, experimental  Vena cava, thrombosis  Veins, iliac  Veins, thrombosis
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号