首页 | 本学科首页   官方微博 | 高级检索  
     


Image guidance to improve reliability and data integrity of transcranial Doppler sonography
Authors:Axel Neulen,Christian Greke,Esther Prokesch,Jochem Kö  nig,Daniel Wertheimer,Alf Giese
Affiliation:1. Department of Neurosurgery, University Medical Center, Johannes Gutenberg-University of Mainz, Langenbeckstrasse 1, 55101 Mainz, Germany;2. Department of Neurosurgery, University of Goettingen, Robert-Koch-Str. 40, 37075 Goettingen, Germany;3. Institute of Medical Biostatistics, Epidemiology and Informatics (IMBEI), University Medical Center, Johannes Gutenberg-University of Mainz, Obere Zahlbacher Straße 69, 55131 Mainz, Germany;4. Department of Neurology, Schoen-Kliniken Eilbeck, Dehnhaide 120, 22081 Hamburg, Germany
Abstract:

Background

Principles and accuracy of image-guided transcranial Doppler (IG TCD) sonography have been published recently. However, it remains open whether combination of image guidance and TCD offers an additional clinical advantage. This study scores the accuracy of conventional TCD examinations and investigates the potential improvement of TCD data integrity and reliability regarding the additional use of IG.

Methods

Conventional TCD was performed by a group of experienced investigators, who were blinded to images of a navigation system tracking the Doppler probe, whereas an independent observer documented the TCD findings, acquired by the investigators, due to saving spatial data of the TCD sample volume using IG for subsequent analysis. In a second set of experiments, image guidance was available to investigators without any previous TCD experience.

Results

The analysis of 3D data of vessels (n = 173) labeled by experienced investigators in conventional TCD, revealed a rate of 37% misinterpreted Doppler signals regarding the target vessel. Correctness of labeling was comparable between the different vascular segments. The rate of correct labeling was higher for right- (69%) than for left-sided vessels (57%). In comparison, by using IG, TCD investigators without any previous TCD experience achieved a significantly lower rate of 10% (n = 39) mislabeled vessels.

Conclusions

Our data suggest, that misinterpretation of the vascular source of the Doppler signal is a common source of errors in conventional TCD. Visualization of the vascular anatomy by image guidance offers improved accuracy and reliability of TCD results and may positively influence the learning curve for inexperienced investigators.
Keywords:Transcranial Doppler sonography   Ultrasound   Image guidance   Neuronavigation   Neurovascular diseases   Vasospasm   Neurointensive care   TCD   ICU   Bone window   Acoustic window   Sonothrombolysis
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏