首页 | 本学科首页   官方微博 | 高级检索  
     


A comparison of flow rates and warming capabilities of the Level 1 and Rapid Infusion System with various-size intravenous catheters
Authors:Barcelona Sandra L  Vilich Fatima  Coté Charles J
Affiliation:Department of Anesthesiology, Northwestern University, The Feinberg School of Medicine, Department of Pediatric Anesthesiology, Children's Memorial Hospital, Chicago, Illinois 60614, USA. s-barcelona@northwestern.edu
Abstract:Cases involving massive blood transfusion may require the use of specialized blood warmers, such as the Level 1 (L-1) (Level 1 Technologies, Inc., Rockland, MA) or the Rapid Infusion System (RIS) (Haemonetics Corp., Braintree, MA). In this in vitro study, we compared the infusion and warming capabilities of the L-1 (model 1000) versus the RIS using pediatric- and adult-sized IV catheters. The time to infuse 2 L of lactated Ringer's solution and the end temperature after infusion through 20-, 18-, 16-, and 14-gauge catheters, and 4-, 5-, 6-, 7-, and 8.5-French catheters using both the L-1 and RIS were measured. The flow rates of both systems were similar for 18- and 20-gauge catheters; however, the flow rates with the RIS were progressively faster than the L-1 as catheter size increased to >18 gauge. The heating capabilities of the RIS were superior to the L-1 for all catheters >or=16 gauge. We conclude that the RIS was superior to the L-1 for both flow rates and warming capacity for all IV catheters >18 gauge, i.e., those used for cases with massive blood loss. The RIS provided no advantage (with regard to heating and flow) when used with typical pediatric-sized catheters. IMPLICATIONS: The rapid infusion system is superior to the Level 1 for warming and flow of crystalloid for IV catheters >18 gauge in vitro. The rapid infusion system provides no advantage with catheters typically used in small children (
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号