头颅柔性线圈改善MRS图像质量的可行性
引用本文:陈义磊,康英杰,谭文莉,詹松华,黄炎文,王辉. 头颅柔性线圈改善MRS图像质量的可行性[J]. 中国中西医结合影像学杂志, 2021, 0(1): 19-23
作者姓名:陈义磊  康英杰  谭文莉  詹松华  黄炎文  王辉
作者单位:;1.上海中医药大学附属曙光医院放射科,上海 201203
基金项目:上海市科学技术委员会科研项目(17441902600,18401970300);2017年度上海中医药大学重大成果培育项目(ZDCG701);上海中医药大学杏林中青年人才培养项目-杏林学者;上海中医药大学附属曙光医院四明学者。
摘    要:目的:探讨头颅柔性12通道线圈(AHC12)改善MRS图像质量的可行性。方法:招募30例健康受试者,采用相同MRI扫描仪及头颅硬质常规16通道线圈(HNC16)、硬质24通道线圈(HC24)、AHC12线圈行MRS成像。采用LCModel软件计算各代谢物与肌酸(Cr)+磷酸肌酸(PCr)的浓度比值,估计标准差值(SD%),并利用联影工作站获得代谢产物峰值、半高全宽值、图像SNR进行客观评价。采用主观评分法评价谱线基线、谱线线型及伪影。结果:图像质量客观评价中,使用HC24及AHC12线圈MRS图像的Cr与Cr+PCr比值、N-乙酰天冬氨酸(NAA)及Cr的SNR均高于HNC16线圈,Cr的SD%值均小于HNC16线圈(均P<0.05);HC24与AHC12间差异无统计学意义(P>0.05)。主观评价中,HC24及AHC12线圈MRS图像的谱线基线、谱线线型评分均高于HNC16线圈(均P<0.05),HC24与AHC12间差异无统计学意义(P>0.05)。结论:在MRS成像中,改良的AHC12线圈可获得与HC24线圈相同的图像质量。

关 键 词:头颅线圈  磁共振波谱学  图像质量

The improved imaging performance in MRS by the flexible head coil with 12 channels for scalp acupuncture
CHEN Yilei,KANG Yingjie,TAN Wenli,ZHAN Songhua,HUANG Yanwen,WANG Hui. The improved imaging performance in MRS by the flexible head coil with 12 channels for scalp acupuncture[J]. Chinese Imaging Journal of Integrated Traditional and Western Medicine, 2021, 0(1): 19-23
Authors:CHEN Yilei  KANG Yingjie  TAN Wenli  ZHAN Songhua  HUANG Yanwen  WANG Hui
Affiliation:(Department of Radiology,Shuguang Hospital Affiliated to Shanghai Traditional Chinese Medicine University,Shanghai 201203,China)
Abstract:Objective:To explore the feasibility that the acupuncture head coil with 12 channels(AHC12)improved the imaging performance of MRS.Methods:Thirty healthy subjects were prospectively recruited and MRS examinations were performed by the same magnetic resonance system and sequence.The different head coils including AHC12,solid head normal coil with 16 channels(HNC16)and solid head coil with 24 channels(HC24)were used randomly.LCModel software was used to process the raw data of MRS.The ratios of metabolites with Cr+PCr and the percent of standard deviation(SD%)of every metabolite were calculated.The peaks of metabolites,full-width at half height(FWHM)and SNR of images were calculated by the uWS-MR workstation.The spectral baseline,the type of spectral baseline and artifact recognition were subjectively evaluated by the Likert 5-point scoring method.Results:In the objective evaluation,the Cr and the SNR of NAA and Cr of HC24 and AHC12 were higher than those of HNC16(all P<0.05),but SD%of Cr of HC24 and AHC12 were lower than that of HNC16(both P<0.05),and there were no statistical differences between AHC12 and HC24(all P>0.05).In the subjective evaluation,the scores of the spectral baseline and the type of spectral line in AHC12 and HC24 were statistically higher than those in HNC16(all P<0.05),and there were no statistical differences between AHC12 and HC24.Conclusion:In the MRS scan,the AHC12 coil can obtain the same clinical image quality as the HC24.
Keywords:Head coil  Magnetic resonance spectroscopy  Image quality