首页 | 本学科首页   官方微博 | 高级检索  
     


Fracture of the femoral component in unicompartmental total knee arthroplasty
Authors:MD H.U. Cameron  MD R.P. Welsh
Affiliation:Orthopaedic and Arthritic Hospital, Toronto, Canada.
Abstract:A fracture of the anterior flange of a femoral component in a unicompartmental knee has been seen. This was thought to be due to lack of bony support leading to cantilever bend. The solution would seem to be to add a metal web to strengthen that area of the femoral component.
Keywords:total knee arthroplasty   unicompartmental knee arthroplasty   knee arthroplasty failure   knee revision   metal failure   tibial subsidence
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏