首页 | 本学科首页   官方微博 | 高级检索  
检索        


Urine-compatible polymer for long-term ureteral stenting
Authors:Cardella  JF; Castaneda-Zuniga  WR; Hunter  DW; Hulbert  JC; Amplatz  K
Abstract:Internal double-J ureteral stents were designed from a urine-compatible polymer (C-Flex), and 35 stents were placed in patients. The overall patency rate for the stents was 80%, with most stent failures occurring before 2 months; the follow-up period ranged from 2 to 16 months, with a mean follow-up for all stents of 5.0 months. Stents were considered patent at last follow-up only if they had been in place for at least 2 months. No migration or fracture of the stents occurred. Physical properties of urine-exposed stents were compared with those of virgin tubing and tubing exposed for 1 year to shelf conditions. Stent patency was optimized by increasing urine flow by increasing the patient's voluntary oral intake, administering prophylactic oral antibiotics, and avoiding placement of stents into grossly bloody or infected collecting systems.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号