The properties of the ultramicrocylindrical ionization chamber for small field used in stereotactic radiosurgery
Authors:Vahc Y W  Chung W K  Park K R  Lee J Y  Lee Y H  Kwon O  Kim S
Institution:Department of Physics, Institute of Basic Medical Science, Yonsei University Wonju College of Medicine, Korea. va23233@wonju.yonsei.ac.kr
Abstract:Accurate dosimetry of small-field photon beams tends to be difficult to perform due to the presence of lateral electronic disequilibrium and steep dose gradients. In stereotactic radiosurgery (SRS), small fields of 6-30 mm in diameter are used. Generally thermoluminescence dosimetry chips, Farmer, Thimble ion chamber, and film dosimetry are not adequate to measure dose in SRS beams. These techniques generally do not provide the required precision due to their energy dependence and/or poor resolution. It is necessary to construct a small, accurate detector with high spatial resolution for the small fields used in SRS. The ultramicrocylindrical ionization chamber (UCIC) with a gold wall of 2.2 mm in diameter and 4.0 mm in length has dual sensitive volumes of air (8.0 mm3) and borosilicate (2.6 mm3) cavity. Reproducibility, linearity, and radiation damage with respect to absorbed dose, beam profile of small beam, and independence of dose rate of the UCIC are tested by the dose measurements in high energy photon (5, 15 MV) and electron (9 MeV) beams. The UCIC with a unique supporting system in the polystyrene phantom is demonstrated to be a suitable detector for the dose measurements in a small beam size.