一种基于多频电阻抗断层成像系统的校准方法研究
引用本文:杨 琳,徐灿华,史学涛,等. 一种基于多频电阻抗断层成像系统的校准方法研究[J]. 医疗卫生装备, 2014, 0(1): 1-5
作者姓名:杨 琳  徐灿华  史学涛  
作者单位:第四军医大学生物医学工程学院,西安710032
基金项目:国家科技支撵计划课题(2011BAl08813);国家自然科学基金课题(51207161);国家自然科学基金重点项目(50937005);国家自然科学基金面上项目(61071033);陕西省自然科学基金课题(2012JQ4012)
摘    要:目的:研究一种简便、有效的多频电阻抗断层成像数据采集系统的幅频和相频校准方法,以提高成像数据的质量.方法:基于系统的正交数字解调方法的特点,提出相频和幅频校正方法;通过电阻网络模型和单通道测量实验给出校正因子的获取方法:开展物理模型实验,对比校准前后的数据质量以及成像结果。结果:物理模型实验表明,校准后的数据质量明显得到改善,成像结果能够清晰地反映成像目标。结论:该校准方法是一种简便、有效的多频电阻抗断层成像系统的幅频和相频校准方法,为未来将多频电阻抗成像技术应用于临床奠定了良好的基础。

关 键 词:多频电阻抗成像数据采集系统  幅频  相频  校准方法

Calibration Method Based on Multi-frequency Electrical Impedance Tomography System
YANG Lin,XU Can-hua,SHI Xue-tao,FU Feng,DAI Meng,XIA Jun-ying,LI Jing,DONG Xiu-zhen. Calibration Method Based on Multi-frequency Electrical Impedance Tomography System[J]. Chinese Medical Equipment Journal, 2014, 0(1): 1-5
Authors:YANG Lin  XU Can-hua  SHI Xue-tao  FU Feng  DAI Meng  XIA Jun-ying  LI Jing  DONG Xiu-zhen
Affiliation:(School of Biomedical Engineering, the Fourth Military Medical University, Xi'an 710032, China)
Abstract:Objective To develop a simple and effective method that calibrates amplitude frequency and phase frequency of multi-frequency EIT data acquisition system, so as the quality of imaging data can be improved to lay a solid foundation for multi-frequency EIT. Methods A calib'ration method of amplitude frequency and phase frequency was proposed based on the system characterization of orthogonal digital demodulation. Subsequently the method to acquire the calibration factor was given by the experiments of resistor network and single-channel measurement. Finally physical phantmn experiments were conducted and data quality as well as image results were contrasted before and after calibration. Results The results of physical phantom experiments showed that after calibration there was a significant improvement of data quality and the imaging target could be clearly reconstructed. Conclusion This method is simple and effective for calibration of amplitude frequency and phase frequency of multi-frequency EIT system.
Keywords:multi-frequency EIT data acquisition system  amplitude frequency  phase fi'equency  calibration method