首页 | 本学科首页   官方微博 | 高级检索  
     


An experimental study of the optimal flip angle for hepatocellular magnetic resonance imaging with Gd-EOB-DTPA
Authors:Chikazawa Sono  Inoue Akira  Tanabe Masahiro  Yamauchi Syuuichi  Nakamura Keiko  Morita Shingo  Uehara Takuya  Mastunaga Naofumi
Affiliation:Department of Radiological Technology, Yamaguchi University Hospital.
Abstract:Gadoxetate Sodium (Gd-EOB-DTPA) is a relatively new contrast agent used for liver examination by magnetic resonance (MR) imaging. The setting of the flip angle (FA) is very important to obtain images with high contrast in hepatocellular MR imaging. The aim of this study was to determine the optimal FA for hepatocellular MR imaging. Optimal FA was estimated using a visual evaluation by observers. We made the visual evaluation with a paired comparison test. When the overall characteristic showed a significant difference, we then evaluated the optimal FA using a yardstick analysis. The visual evaluation by observers indicated that the FA was best at 22 degrees for the overall characteristic, the contrast characteristic, and the sharpness characteristic, and for the noise characteristic, the best FA was 18 degrees. All characteristics showed a significant difference between 18 and 22 degrees. Based on our results, the most important factor was the contrast characteristic for the visual evaluation, and we concluded that the optimal FA in hepatocelluar MR imaging was 22 degrees.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号