首页 | 本学科首页   官方微博 | 高级检索  
检索        


Fast 31P chemical shift imaging using SSFP methods.
Authors:O Speck  K Scheffler  J Hennig
Institution:Department of Diagnostic Radiology, Section Medical Physics, University Medical Center, Freiburg, Germany. ospeck@ukl.uni-freiburg.de
Abstract:Steady-state free precession (SSFP) methods have been very successful due to their high signal and short imaging times. These properties make them good candidates for applications that intrinsically suffer from low signal such as low gamma nuclei imaging. A new chemical shift imaging (CSI) technique based on the SSFP signal formation has been implemented and applied to (31)P. The signal properties of the SSFP CSI method have been evaluated and the steady-state signal of (31)P has been measured in human muscles. Due to the T(2) and T(1) signal dependence of SSFP, the steady-state signal mainly consists of phosphocreatine (PCr). The technique allows fast CSI acquisitions with high SNR of the PCr signal. The SNR gain for PCr over a FLASH-based CSI method is approx. 4-5. Fast in vivo CSI of human muscle with subcentimeter resolution and high SNR is demonstrated at 2 T.
Keywords:31P‐imaging  SSFP  fast imaging  chemical shift imaging  phosphocreatine
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号