Comparison of seven different reagents of peroxidase method for small and dense low density lipoprotein cholesterol (sdLDL‐C) measurement
Authors:Fan Xuesong  Wang Enshi  He Jianxun  Zhang Lei  Zeng Xiaoli  Yuan Hui
Institution:1. Department of Clinical Laboratory Center, Beijing AnZhen Hospital, Capital Medical University, Beijing Institute of Heart, Lung and Blood Vessel Diseases, Beijing China ; 2. Department of Cardiac Surgery, Fuwai Hospital, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing China
Abstract:BackgroundWe validated the performance of seven different reagents of peroxidase method for sdLDL‐C in two automatic analyzers that are common in Chinese laboratories.MethodsSeven commercially available sdLDL‐C assays were analyzed with the Beckman AU5400 and Mindray BS2000 automatic analyzers. A total of 336 blood samples were collected and the reference interval was also validated in 298 apparently healthy individuals. Serum samples were used for method comparison of precision, recovery, lower limit of detection, comparison and concurrence analysis, as well as reference interval for the Mindray reagent.ResultsThe repeatability CV% of the seven sdLDL‐C assays were 0.81%~3.66% for Mindray BS2000 and 0.76%~3.91% for Beckman AU5400, while Total CVs for Mindray BS2000 sdLDL‐C assay were 1.34%~4.81%, and that of Beckman AU5400 were 2.25%~10.33%. The measured recovery rates of sdLDL‐C assays were within the allowable ±10% deviation range. There was no obvious difference between the reagents in the lower limit detection. There was a difference between the validation results of the reference range and the manufacturer''s.BSBE, Mindray, and Dongou had a high degree of association with DENKA SEIKEN on Mindray BS2000, while BSBE, Mindray, Dongou and Merit Choice had a high degree of association with DENKA SEIKEN on Beckman AU5400. Passing–Bablok regression showed excellent linear correlation between BSBE and Mindray and DENKA SEIKEN and on Beckman AU5400.ConclusionsOur results indicate that the basic performance can meet the testing requirements, but the comparability between them is still insufficient.
Keywords:method comparison  peroxidase method  reference interval  small and dense low density lipoprotein cholesterol
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号