Dose response evaluation of a low-density normoxic polymer gel dosimeter using MRI |
| |
Authors: | Haraldsson P Karlsson A Wieslander E Gustavsson H Bäck S A J |
| |
Affiliation: | Medical Radiation Physics, Department of Clinical Sciences, Lund University, Malm? University Hospital, SE-205 02 Malm?, Sweden. |
| |
Abstract: | A low-density (approximately 0.6 g cm(-3)) normoxic polymer gel, containing the antioxidant tetrakis (hydroxymethyl) phosponium (THP), has been investigated with respect to basic absorbed dose response characteristics. The low density was obtained by mixing the gel with expanded polystyrene spheres. The depth dose data for 6 and 18 MV photons were compared with Monte Carlo calculations. A large volume phantom was irradiated in order to study the 3D dose distribution from a 6 MV field. Evaluation of the gel was carried out using magnetic resonance imaging. An approximately linear response was obtained for 1/T2 versus dose in the dose range of 2 to 8 Gy. A small decrease in the dose response was observed for increasing concentrations of THP. A good agreement between measured and Monte Carlo calculated data was obtained, both for test tubes and the larger 3D phantom. It was shown that a normoxic polymer gel with a reduced density could be obtained by adding expanded polystyrene spheres. In order to get reliable results, it is very important to have a uniform distribution of the gel and expanded polystyrene spheres in the phantom volume. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|