Black-on-clear piggyback technique for a black occlusive intraocular device in intractable diplopia |
| |
Authors: | Byard Stephen D Lee Richard M H Lam Fook Chang Simpson Andrew R H Liu Christopher S C |
| |
Affiliation: | From Sussex Eye Hospital (Byard, Lee, Lam, Simpson, Liu), Brighton, and Tongdean Eye Clinic (Liu), Hove, United Kingdom |
| |
Abstract: | Black occlusive intraocular devices have been used successfully for intractable binocular diplopia. We describe a novel technique of implanting both a black occlusive device and a clear poly(methyl methacrylate) intraocular lens (IOL) in the capsular bag during phacoemulsification surgery. If the need should arise at a later date, this approach will allow safer and easier explantation of the black occlusive device, avoiding the need for IOL exchange. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|