A novel patient-specific instrument design can deliver robotic level accuracy in unicompartmental knee arthroplasty |
| |
Authors: | Gareth G. Jones Susannah Clarke Simon Harris Martin Jaere Thunayan Aldalmani Patrick de Klee Justin P. Cobb |
| |
Affiliation: | 2. Royal College of Surgeons in Ireland, 123 St. Stephen''s Green, Dublin 2, Ireland;3. Embody Orthopaedic, MSk Lab, 7th Floor Lab Block, Charing Cross Hospital, London, W6 8RF, UK |
| |
Abstract: | BackgroundA previous randomised controlled trial (RCT) by our group found that robotic assisted unicompartmental knee arthroplasty (UKA) surgery was significantly more accurate than conventional instrumentation. The aim of this study was to determine whether a low-cost novel PSI design could deliver the same level of accuracy as the robot in the same time efficient manner as conventional instruments.MethodsThirty patients undergoing medial UKA took part. Tibial component position was planned using a low dose CT-scan, and compared to a day 1 postoperative CT-scan to determine the difference between the planned and achieved positions. Operations were performed by one expert surgeon using PSI (Embody, London, UK).ResultsThe mean absolute difference between planned and achieved tibial implant positions using PSI was 2.0° (SD 1.0°) in the coronal plane, 1.8° (SD 1.5) in the sagittal plane, and 4.5° (SD 3.3) in the axial plane. These results were not significantly different to the 13 historical robotic cases (mean difference 0.5°, 0.5°, and 1.7°, p = 0.1907, 0.2867 and 0.1049 respectively). PSI mean operating time was on average 62 min shorter than the robotic group (p < 0.0001) and 40 min shorter than the conventional instrument group (p < 0.0001). No complications were reported.ConclusionsIn conclusion, this clinical trial demonstrates that for tibial component positioning in UKA, a novel design PSI guide in the hands of an expert surgeon, can safely deliver comparable accuracy to a robotic system, whilst being significantly faster than conventional instruments.NIHR Clinical Research Network Reference: 16100. |
| |
Keywords: | Corresponding author. Unicompartmental knee arthroplasty Partial knee replacement Patient specific instrumentation PSI Robotic surgery Robot-assisted surgery |
本文献已被 ScienceDirect 等数据库收录! |
|