Clinical Evaluation of a Dual Sensor, Rate Responsive Pacemaker
Authors:SANJAY SINHA  RICHARD J. SCHILLING  GERRY C. KAYE  JOHN L. CAPLIN
Affiliation:Department of Cardiology, Castle Hill Hospital, Cottingham, United Kingdom;Department of Cardiology, Hull Royal Infirmary, Hull, United Kingdom
Abstract:Dual sensor pacemakers should respond more appropriately during differing exercise modes than a single sensor device. The Topaz™ models 515 (QT and activity count [ACT] sensing) pacemaker shows appropriate rate response during treadmill exercise testing. We postulated that adjustments to relative sensor contribution should allow fine tuning of the onset of rate response. Eleven patients with this pacemaker were studied. Three standard exercise tests were performed with adjustment of sensor blending and activity threshold between each one. We also assessed the response to isometric exercise and a false positive activity signal. Results : Times to 100 ppm (3.7 ± 1.3, 4.4 ± 2.0, 5.3 ±1.5 mins), times to peak rate (6.1 ± 1.6, 5.6 ± 1.4, 6.5 ± 1.3 mins) and accelerations to peak (9.0 ± 2.4, 9.2 ± 5.3, 7.7 ± 2.8 ppm/min) were measured in all three different sensor settings (QT = ACT, QT
Keywords:dual sensor    rate response    pacemaker    isometric exercise