首页 | 本学科首页   官方微博 | 高级检索  
     


Quality assurance of brachytherapy afterloaders using a multi-slit phantom
Authors:Kohr Peter  Siebert Frank-André
Affiliation:University Clinic of Schleswig-Holstein, Campus Kiel, Clinic of Radiotherapy, Arnold-Heller-Str. 9, 24105 Kiel, Germany.
Abstract:Brachytherapy is a very successful treatment option for many different tumors. Consequently the quality assurance (QA) of afterloaders is of special interest. For many quality checks verification films are essential. Because currently in many clinics the film development units are replaced by computed radiography systems, the question arises of how to perform QA for afterloaders without conventional radiography films. In this note the use of a new designed multi-slit phantom demonstrates the possibility of performing QA checks with high sensitivity computed radiography systems. Because the presented phantom allows a method for imaging the radiation source it might be useful in the development of further QA techniques.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏