GuidezillaTM延长导管在复杂冠状动脉介入治疗中的应用
引用本文:袁思敏,赵秀峰,信栓力,常超,刘丽军,韩丽英.GuidezillaTM延长导管在复杂冠状动脉介入治疗中的应用[J].临床荟萃,2019,34(8):720.
作者姓名:袁思敏  赵秀峰  信栓力  常超  刘丽军  韩丽英
作者单位:邯郸市第一医院 心内一科,河北 邯郸 056000
摘    要:目的 对应用GuidezillaTM延长导管在复杂经皮冠状动脉介入治疗(PCI)中的有效性及安全性进行评估, 并总结应用GuidezillaTM临床经验。方法 2017年1月至2018年12月邯郸市第一医院PCI手术中应用因球囊或支架难以通过,补救性应用GuidezillaTM辅助球囊或支架通过病变进行PCI手术患者53例。收集术中穿刺入路、病变性质、靶血管开口变异、钙化迂曲、特殊操作及GuidezillaTM延长导管深插冠状动脉长度等数据,观察应用该器械的安全性和有效性。结果 应用GuidezillaTM辅助球囊和支架进行手术患者53例,其中成功植入支架49例(92.5%)。经右桡动脉途径45例、经右股动脉进行手术8例,靶血管开口异常4例,靶血管开口存在病变22例,慢性完全闭塞病变19例,钙化病变19例,迂曲病变9例,钙化迂曲病变24例,辅助旋磨磨头达到病变1例,GuidezillaTM深插冠状动脉深度(19.38±12.66) mm。术中术后无严重并发症,无死亡病例。结论 在复杂PCI治疗中,补救性应用GuidezillaTM辅助介入治疗安全有效,手术成功率高,安全有效,并发症少。

关 键 词:血管成形术     气囊     冠状动脉  GuidezillaTM延长导管  复杂病变  

Case series experience of GuidezillaTM extension catheter
Yuan Simin,Zhao Xiufeng,Xin Shuanli,Chang Chao,Liu Lijun,Han Liying.Case series experience of GuidezillaTM extension catheter[J].Clinical Focus,2019,34(8):720.
Authors:Yuan Simin  Zhao Xiufeng  Xin Shuanli  Chang Chao  Liu Lijun  Han Liying
Institution:First  Department of  Cardiology,Handan First Hospital,Handan 056000, China
Abstract:Objective To summarize reportson experience using the GuidezillaTM extension catheter for complex coronary lesions at the documented institution and evaluation of its safety and efficacy. Methods We retrospectively collected 53 cases from the Cardiology Department of Handan First Hospital from January of 2017 to December of 2018 who had received percutaneous coronary intervention (PCI). GuidezillaTM extension catheter was used during PCI Since the balloon or stent cannot pass leision.The clinical, angiographic and procedural data of all 53 procedures were collected. the safety and efficacy of this novel equipment was evaluated.In addition,recorde the successful rate and the incidence of complications. the adverse cardiovascular events were also recorded during follow up. Results 53 patients were received with GuidezillaTM extension assisted balloon and stent to achieve coronary artery lesions for PCI. The mean depth of intubation was 19.38±12.66 mm.Stent implantation was successful in 49 out of 53 procedures (92.5%). 45 cases via radial artery, 8 cases via arteria femoralis, 4 cases of target abnormal openings, 22 cases of target vessel narrow openings, 19 cases of chronic, totally occluded (CTO) 9 cases circuity pathological changes, 19 cases of calcified lesions, 9 cases of tortuosity lesions and 24 case of tortuosity and calcified lesions.1cases of rotational atherectomy. None of the patients experienced coronary dissection, thrombogenesis air embolism, or other major complications during the procedure.Conclusion The use of GuidezillaTMguide extension catheter to support extension catheter assisted balloon and stent to reach the lesion site of complex coronary lesion is safe and effective and has a higher success rate and less complications.
Keywords:angioplasty     balloon     coronary  GuidezillaTMextensioncatheter  complex coronary lesion  