In vitro oxyhaemoglobin saturation measurements in haemoglobin solutions using fibreoptic pulmonary artery catheters
Authors:KONG, C. S.   RYDER, I. G.   KAHN, R.   GREGORY, L.   MACKENZIE, C. F.
Affiliation:Department of Anaesthesiology, University of Maryland School of Medicine 22 South Greene Street, Baltimore, Maryland 21201, USA
Department of Medical and Research Technology, University of Maryland School of Medicine Allied Heath Building, Penn Street, Baltimore, Maryland 21201, USA
Abstract:We compared in vitro oxyhaemoglobin saturations using two pulmonaryartery catheters (catheter So2) with oxyhaemoglobin saturations(So2) measured by the IL282 co-oximeter and derived partialoxyhaemoglobin saturations (partial So2) at different oxygentensions (Po2) in six solutions: whole blood, 50:50 mixtureof whole blood and Plasmalyte A (haemodiluted blood), 50:50mixture of whole blood and 8% pyridoxylated haemoglobin-polyoxyethylene(PHP) conjugate (WB-PHP), 75:25 mixture of 8% PHP and PlasmalyteA solution (PHP66), 50:50 mixture of 8% PHP and Plasmalyte Asolution (PHP44) and stroma-free haemoglobin solution (SFH).Calculated P50 values (Po2, vs So2) were 3.79, 3.58, 3.49, 3.15,3.04 and 2.07 kPa, respectively. However, if partial So2 wasused the curves were shifted to the left, reducing P50. CatheterSo2 correlated well with So in whole blood (r2 > 0.99 forboth catheters), haemodiluted blood (r2 > 0.98 for both catheters)and WB-PHP solution (r2 = 0.94 for both catheters). In PH P44(r2 = 0.64 and r2 = 0.57), PHP P66 (r2 = 0.40 for the Oximetrixand r2 = 0.25 for the Edwards catheter) and SFH solutions (r2= 0.33 for the Oximetrix and r2 = 0.22 for the Edwards catheter)both catheters performed poorly. We conclude that mixed venousoxyhaemoglobin saturations measured by oximetric pulmonary arterycatheters are inaccurate in the presence of haemoglobin solutions.For accuracy a multi-wavelength co-oximeter should be used ifblood containing PHP or SFH is to be analysed. {dagger}Present address: Department of Anaesthetics, Bristol Royal Infirmary,Bristol, UK
Keywords:
本文献已被 Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号