首页 | 本学科首页   官方微博 | 高级检索  
     


Electronic and disposable patient-controlled analgesia systems
Authors:S. L. Robinson  D. J. Rowbotham  M. Mushambi
Affiliation:University Department of Anaesthesia, Leicester Royal Infirmary.
Abstract:We have compared the performance of the Baxter disposable with the Graseby electronic patient-controlled analgesia system in 30 patients following major gynaecological surgery. Patients were allocated randomly to receive analgesia via the Baxter or Graseby device for postoperative pain relief. There were no significant differences between the two groups with regard to postoperative pain relief or sedation as measured by visual analogue scale. Requirements for antiemetic drugs and patient acceptability were similar. Mean (SEM) morphine demanded over 30 h was 35.7 (6.6) in the Graseby group and 35.1 (8.5) in the Baxter group.
Keywords:Analgesia    postoperative    Equipment    patient-controlled analgesia systems