Radiological placement of the AshSplit haemodialysis catheter: a prospective analysis of outcome and complications. |
| |
Authors: | Fiona Ewing Dilip Patel Anne Petherick Robin Winney Kieran McBride |
| |
Affiliation: | Department of Radiology, Royal Infirmary of Edinburgh, Edinburgh, UK. |
| |
Abstract: | BACKGROUND: The AshSplit catheter has recently been introduced as an alternative permanent tunnelled haemodialysis catheter, combining ease of insertion with good long-term patency and flow rates. METHODS: Data were collected prospectively on all the long-term tunnelled haemodialysis (AshSplit) catheters inserted radiologically between January 1998 and March 1999. Information was obtained regarding the initial insertion, ongoing catheter function and re-intervention up to September 1999. RESULTS: A total of 118 catheters were inserted in 88 patients (50 male), median (range) age 64 (20-86) years. Ultrasound guidance was used routinely and the right internal jugular vein was used in 80 (68%) cases. Initial complications occurred in 14 (11.9%) cases, which included local haemorrhage, carotid artery puncture, and air embolism. Infection occurred in 34% of catheters (2.4/1000 catheter days). Line thrombosis was documented in 20% (1.2/1000 catheter days). Satisfactory mean urea reduction ratio (URR) of 63 was obtained for all catheters. There were 47 re-interventions, mainly for fibrin sheath stripping (34) and/or thrombectomy (25). Total catheter duration was 21600 days with a 1 month cumulative survival of 87% (Kaplan-Meier probability 85%). At the end of the study, 20 (17%) catheters were still functioning, 39 (33%) had been removed electively, and 22 (18%) patients had died with a functioning catheter in situ. Catheter infection was implicated in four deaths. CONCLUSIONS: Radiological insertion of the AshSplit catheter is well tolerated, providing reliable short- and long-term dialysis access. Radiology also has a role in maintaining patency. As with all tunnelled catheters, infection remains a problem. |
| |
Keywords: | access catheters haemodialysis long-term split tunnelled |
本文献已被 Oxford 等数据库收录! |
|